Gilson F100

Gilson Pipetman F100 Model F Fixed-Volume Pipette, 100µl Volume Range Instruction Manual

1. Introdución e finalview

The Gilson Pipetman F100 is a fixed-volume, air-displacement pipette designed for precise liquid handling in laboratory settings. This model is specifically calibrated for a 100µl volume range, ensuring consistent and reliable results for routine testing, clinical diagnostics, and quality control applications. Its fixed-volume mechanism eliminates the risk of volume-selection errors during standard operation.

While primarily fixed, the pipette's volume can be adjusted by qualified users in a laboratory environment to accommodate for specific fluid properties, such as dense or viscous liquids. Once reset, the new volume remains fixed and cannot be accidentally altered during use.

Gilson Pipetman F100 Fixed-Volume Pipette, showing the main body, plunger, and tip ejector, with a disposable tip next to it.

Figura 1: The Gilson Pipetman F100 fixed-volume pipette. This image displays the blue main body with the 'Pipetman' logo, a yellow plunger mechanism visible at the top, and a white tip ejector rod extending downwards. A clear disposable pipette tip is shown separately, ready for attachment.

2. Configuración

  1. Desembalaxe: Carefully remove the Pipetman F100 from its packaging. Inspect the pipette for any visible damage that may have occurred during transit.
  2. Inspección inicial: Ensure all components are present and securely attached. The pipette should feel solid and free of loose parts.
  3. Adxunto de consellos: Select a high-quality, compatible pipette tip. Firmly press the pipette shaft into the opening of the tip with a slight twisting motion until a secure seal is formed. Avoid excessive force, which can damage the tip cone or affect accuracy.
  4. Volume Verification (Optional): Although fixed, it is recommended to perform an initial gravimetric check with distilled water to confirm the pipette's accuracy, especially if it is new or has been stored for an extended period. Refer to standard laboratory procedures for gravimetric calibration.

3. Instrucións de funcionamento

The Gilson Pipetman F100 is designed for straightforward operation due to its fixed-volume nature. Follow these steps for accurate liquid transfer:

  1. Attach a Tip: Securely attach a new, clean pipette tip to the tip cone as described in the Setup section.
  2. Pre-wetting (Recommended): For optimal accuracy, especially with volatile liquids, aspirate and dispense the liquid once or twice before transferring the actual sample. This saturates the air space in the tip with vapor, reducing evaporative loss during the actual transfer.
  3. Aspiración:
    • Depress the plunger smoothly to the first stop.
    • Immerse the tip into the liquid to a depth of 2-3 mm.
    • Slowly release the plunger to allow the liquid to be drawn into the tip. Wait 1-2 seconds after the plunger reaches its resting position to ensure the full volume is aspirated.
    • Withdraw the tip from the liquid, touching it against the edge of the reservoir to remove any excess droplets.
  4. Distribución:
    • Coloque a punta contra a parede interior do recipiente receptor cun lixeiro ángulo.
    • Smoothly depress the plunger to the first stop, expelling the majority of the liquid.
    • Continue depressing the plunger to the second stop (blow-out) to ensure complete delivery of the remaining liquid.
    • While holding the plunger at the second stop, withdraw the pipette from the vessel.
  5. Expulsión da punta: Press the tip ejector button to discard the used tip into an appropriate waste container. Never reuse tips.

4. Mantemento

Proper maintenance ensures the longevity and accuracy of your Gilson Pipetman F100.

5 Solución de problemas

If you encounter issues with your Pipetman F100, consider the following common problems and solutions:

ProblemaCausa posibleSolución
Inaccurate or inconsistent dispensing
  • Loose or incompatible tip
  • Damaged tip cone
  • Contamination inside the pipette
  • Improper pipetting technique (e.g., too fast aspiration/dispensing)
  • Temperature differences between pipette, liquid, and environment
  • Ensure tip is securely attached and compatible.
  • Inspect tip cone for cracks or damage; replace if necessary.
  • Perform thorough cleaning and decontamination.
  • Review operating instructions for correct technique.
  • Allow pipette and liquids to equilibrate to room temperature.
  • Perform a gravimetric calibration check.
Liquid dripping from tip
  • Loose or incompatible tip
  • Damaged tip cone or seal
  • Contamination affecting seal
  • Re-attach tip firmly or try a new tip.
  • Inspect tip cone for damage.
  • Clean the tip cone and shaft.
  • If problem persists, contact Gilson support.
Plunger movement is stiff or sticky
  • Falta de lubricación
  • Contamination in the plunger mechanism
  • Consult the full manufacturer's manual for lubrication instructions.
  • Disassemble and clean the plunger assembly if comfortable, or send for professional servicing.

For issues not listed here or if problems persist, refer to the comprehensive Gilson Pipetman F100 user manual or contact Gilson technical support.

6. Especificacións

CaracterísticaDetalle
ModeloPipetman F100
Rango de volume100 µL (Fixed)
Tipo de pipetaAir-displacement, Fixed-Volume
FabricanteGilson
Dimensións do paquete12.5 x 4.3 x 1.5 polgadas (31.75 x 10.92 x 3.81 cm)
Peso do paquete8 onzas (0.5 libras)
ASINB008TPWFLA

7. Garantía e soporte

For detailed warranty information, service, or technical support regarding your Gilson Pipetman F100, please refer to the official Gilson website or contact Gilson customer service directly. Ensure you have your product model number (F100) and any purchase details available when seeking support.

Visit Gilson's Official Websitio

Documentos relacionados - F100

Preview PIPETMAN® L User's Guide: Precision Pipetting for Laboratories
Comprehensive user guide for the Gilson PIPETMAN® L series of pipettes, detailing specifications, operation, maintenance, and troubleshooting for single-channel and multichannel models.
Preview Guía rápida da pipeta Gilson PIPETMAN Classic: especificacións e uso
Unha guía concisa da pipeta Gilson PIPETMAN Classic, que abrangue as súas características, especificacións, erros máximos permitidos, axuste de volume, axuste da punta, técnicas de pipeteado, mantemento e información sobre a garantía.
Preview Guía rápida de Gilson PIPETMAN G: funcionamento e especificacións da pipeta
Unha guía concisa da pipeta de desprazamento de aire Gilson PIPETMAN G, que abrangue as súas características, pezas, especificacións, axuste de volume, axuste da punta, mantemento e garantía.
Preview Guía rápida da pipeta Gilson PIPETMAN Classic: características, especificacións e uso
Unha guía concisa da pipeta monocanal Gilson PIPETMAN Classic, que abrangue as súas características, especificacións, erros máximos permitidos, axuste de volume, axuste da punta, técnicas de pipeteado, mantemento e información sobre a garantía.
Preview Guía rápida da pipeta Gilson PIPETMAN
Unha guía concisa para o uso das pipetas Gilson PIPETMAN, que abrangue as especificacións, o funcionamento, o mantemento e a información sobre a garantía para modelos monocanal e multicanal.
Preview Gilson PIPETMAN Replacement Parts Guide
A comprehensive guide to replacement parts for Gilson PIPETMAN pipettes, including PIPETMAN Classic, Fixed, Neo, G, L, M, Microman, and Distriman models. Find part numbers and service kits for various pipette capacities.