1. Introdución
This manual provides instructions for the proper setup, operation, and maintenance of the Apera Instruments MP511 Benchtop pH Meter. The MP511 is designed for accurate pH measurements with a range of -2.00 to 19.99 pH and an accuracy of ±0.01 pH. It features automatic temperature compensation (ATC), auto-calibration with buffer recognition, and advanced digital filtering technology for improved response time and accuracy. The meter also includes GLP standard data management, storage for 500 data sets, and RS232 data output.
2. Contido do paquete
Verifique que estean presentes todos os compoñentes ao desembalar:
- MP511 Benchtop pH Meter Unit
- 3-in-1 Combination pH Electrode
- Flexible Electrode Holder
- RS232 Software (typically on a CD or available for download)
- Adaptador de alimentación
- pH Buffer Solutions (e.g., pH 4.00, 7.00, 10.01)
- Manual de instrucións
3. Configuración
- Assemble the Electrode Holder: Attach the flexible electrode holder to the meter base. Ensure it is securely fastened to provide stable support for the electrode.
- Connect the Electrode: Plug the BNC connector of the 3-in-1 combination pH electrode into the pH/mV input port on the back of the meter. If a separate temperature probe is used, connect it to the TEMP port.
- Conectar enerxía: Plug the DC 9V power adapter into the DC 9V port on the back of the meter and then into a standard electrical outlet.
- Colocación inicial: Place the electrode in its storage solution or a beaker of distilled water when not in use.

Figura 3.1: The Apera Instruments MP511 Benchtop pH Meter with its electrode immersed in a sample solution within a beaker, demonstrating a typical setup for pH measurement. The flexible electrode holder is visible, supporting the electrode.
4. Funcionamento
- Encendido/apagado: Preme o ON/OFF botón para acender ou apagar o medidor.
- Selección de modo: Preme o MODO button to cycle through different measurement modes (pH, mV, temperature).
- Selección de unidades: Preme o UNIDADE button to change temperature units between °C and °F.
- Data Storage/Recall: Usa o M+/RM button to store current readings or recall previously stored data.
- Confirm/Enter: O ENTRAR button is used to confirm selections or enter calibration mode.

Figura 4.1: Un detallado view of the MP511 meter's LCD display and control panel, showing the CAL, ON/OFF, M+/RM, MODE, ENTER, and UNIT buttons. The display shows pH and temperature readings.
5. Calibración
Regular calibration is essential for accurate pH measurements. The MP511 features automatic calibration with buffer recognition.
- Prepare Buffers: Ensure you have fresh pH buffer solutions (e.g., pH 4.00, 7.00, 10.01) at room temperature.
- Rinse Electrode: Rinse the electrode thoroughly with distilled water and gently blot dry with a lint-free tissue.
- Entrar no modo de calibración: With the meter powered on and in pH mode, press the CAL button. The display will indicate calibration mode.
- First Buffer Calibration: Immerse the electrode into the first buffer solution (e.g., pH 7.00). The meter will automatically recognize the buffer and stabilize. Press ENTRAR para confirmar.
- Subsequent Buffer Calibration: Rinse the electrode again. Immerse it into the next buffer solution (e.g., pH 4.00 or pH 10.01). The meter will recognize and stabilize. Press ENTRAR to confirm. Repeat for additional buffer points as needed.
- Saír da calibración: After completing all desired calibration points, the meter will automatically return to measurement mode or prompt to save calibration.
6. Modos de medición
The MP511 offers three distinct measurement modes:
- Stable Measurement Display Mode: Displays the pH reading once it has stabilized. This is the standard mode for most measurements.
- Timing Measuring Mode: Allows for measurements at set intervals, useful for monitoring changes over time.
- Modo de medición continua: Provides real-time, continuous pH readings, suitable for dynamic processes.
Refer to the detailed instructions in the full manual for specific steps on how to select and configure each measurement mode.
7. Xestión de datos
The MP511 incorporates GLP (Good Laboratory Practice) standard data management features:
- Almacenamento de datos: The meter can store up to 500 sets of measurement data.
- Métodos de almacenamento: Data can be stored manually or automatically at user-defined timing intervals.
- Saída de datos: An RS232 port is provided for connecting the meter to a computer to transfer stored data using the included software.

Figura 7.1: The rear panel of the MP511 meter, displaying the REF, pH/mV, TEMP, RS232, and DC 9V input ports. The RS232 port facilitates data transfer to a computer.
8. Mantemento
Proper maintenance ensures the longevity and accuracy of your pH meter.
- Coidado dos eléctrodos:
- Always keep the electrode bulb immersed in storage solution when not in use. Never store it dry.
- Rinse the electrode with distilled water before and after each measurement.
- Clean the electrode regularly according to the electrode's specific cleaning instructions, especially if readings become sluggish or inaccurate.
- Meter Cleaning: The MP511 meter unit meets IP54 waterproof and dustproof standards. Clean the exterior with a damp pano. Evite o uso de produtos químicos agresivos ou materiais abrasivos.
- Fonte de enerxía: The meter is battery-powered (though also uses a DC 9V adapter). Ensure the power adapter is used correctly. If using internal batteries, replace them as needed.
9 Solución de problemas
| Problema | Causa posible | Solución |
|---|---|---|
| Lecturas inestables | Dirty electrode; air bubbles; insufficient sample volume; electromagnetic interference | Clean electrode; gently tap electrode to remove bubbles; ensure adequate sample; move meter away from interference sources |
| Erro de calibración | Expired or contaminated buffer solutions; dirty electrode; incorrect buffer sequence | Use fresh buffer solutions; clean electrode; ensure correct buffer order (e.g., 7.00 first) |
| Sen problema de pantalla/alimentación | Power adapter not connected; faulty power adapter; internal battery depletion | Check power connection; try a different outlet; replace internal batteries if applicable |
| Tempo de resposta lento | Electrode dry or contaminated; old electrode | Rehydrate electrode in storage solution; clean electrode; consider electrode replacement |
10. Especificacións
| Parámetro | Valor |
|---|---|
| Número de modelo | AI511 (MP511) |
| Rango de pH | -2.00 a 19.99 pH |
| Precisión do pH | ± 0.01 pH |
| Compensación de temperatura | Automatic (ATC) |
| Calibración | Auto Calibration with buffer recognition |
| Almacenamento de datos | 500 data sets (manual/auto timing storage) |
| Conectividade | RS232 |
| Tipo de visualización | LCD con retroiluminación azul |
| Clasificación de protección de entrada (IP). | IP54 (resistente á auga e ao po) |
| Dimensións do produto | 19.05 x 16 x 7.11 cm |
| Peso do produto | 878.84 g |
| Fonte de enerxía | Battery Powered (with DC 9V adapter) |
| Fabricante | Shanghai Sanxin Instrumentation |
11. Garantía e soporte
For warranty information, technical support, or service inquiries, please contact Apera Instruments directly through their official websitio web ou canles de atención ao cliente. Garda o teu recibo de compra como proba de compra.