1. Introdución

This manual provides comprehensive instructions for the proper preparation, use, and maintenance of the OHAUS pH Buffer Powder Set. This set includes powders for pH 4.01, 7.00, and 10.01, designed for accurate calibration of pH meters in laboratory and industrial settings. Adhering to these instructions will ensure optimal performance and longevity of your pH measurement equipment.

2. Produto rematadoview

The OHAUS 83033971 pH Buffer Powder Set consists of pre-measured powder sachets for preparing standard pH buffer solutions. Each sachet is formulated to produce a precise pH value when dissolved in a specified volume of deionized water. The set includes:

OHAUS pH Buffer Powder Set packaging

Figure 2.1: Packaging of the OHAUS pH Buffer Powder Set, showing the individual sachets.

OHAUS brand logo

Figure 2.2: The OHAUS brand logo, representing the manufacturer of the pH buffer powder set.

3. Setup: Preparing Buffer Solutions

To ensure accurate pH calibration, it is crucial to prepare the buffer solutions correctly. Use high-quality deionized or distilled water for preparation.

  1. Recollida de materiais: You will need one sachet of the desired pH buffer powder, a clean 250 mL volumetric flask (or other appropriate volume as specified on the sachet), a stirring rod, and deionized/distilled water.
  2. Open Sachet: Carefully open one sachet of the pH buffer powder. Avoid spilling the contents.
  3. Transfer Powder: Transfer the entire contents of the sachet into the clean volumetric flask.
  4. Engadir auga: Add a small amount of deionized water to the flask, just enough to dissolve the powder. Stir gently with the stirring rod until the powder is completely dissolved.
  5. Dilute to Volume: Carefully add deionized water to the flask until the bottom of the meniscus aligns with the calibration mark on the neck of the volumetric flask. Ensure the solution is at room temperature (25°C) for accurate volume measurement.
  6. Mesturar ben: Stopper the flask and invert it several times to ensure the solution is thoroughly mixed.
  7. Label Solution: Label the prepared buffer solution with its pH value, preparation date, and expiration date (typically 30 days after preparation or as specified by the sachet).

Repeat this process for each pH buffer powder (4.01, 7.00, 10.01) you intend to use.

4. Operating Instructions: pH Meter Calibration

Calibrating your pH meter with fresh buffer solutions is essential for accurate measurements. Always refer to your specific pH meter's instruction manual for detailed calibration procedures, as steps may vary slightly.

  1. Prepare pH Meter: Ensure your pH meter and electrode are clean and functioning correctly. Rinse the electrode with deionized water and gently blot dry with a lint-free tissue.
  2. Select Calibration Mode: Place the pH meter in its calibration mode.
  3. First Buffer (pH 7.00): Immerse the pH electrode into the pH 7.00 buffer solution. Ensure the electrode's sensing bulb and junction are fully submerged. Allow the reading to stabilize. Confirm the calibration point on your meter.
  4. Rinse Electrode: Remove the electrode from the pH 7.00 buffer, rinse thoroughly with deionized water, and gently blot dry.
  5. Second Buffer (pH 4.01 or 10.01): Immerse the electrode into either the pH 4.01 or pH 10.01 buffer solution, depending on your expected sample pH range. For acidic samples, use pH 4.01. For alkaline samples, use pH 10.01. Allow the reading to stabilize and confirm the second calibration point.
  6. Third Buffer (Optional): For enhanced accuracy, especially across a wide pH range, perform a three-point calibration using all three buffers (4.01, 7.00, and 10.01). Rinse the electrode between each buffer.
  7. Saír do modo de calibración: Once calibration is complete, exit the calibration mode on your pH meter. Your meter is now ready for sample medidas.

Always calibrate your pH meter before each measurement session or if significant temperature changes occur.

5. Mantemento e almacenamento

Proper storage of both the powder sachets and prepared solutions is vital for maintaining their accuracy and shelf life.

5.1 Powder Sachets

5.2 Prepared Buffer Solutions

6 Solución de problemas

If you encounter issues during calibration or measurement, consider the following common problems and solutions:

ProblemaCausa posibleSolución
Inaccurate pH readings after calibration
  • Expired or contaminated buffer solutions
  • Incorrect buffer preparation
  • Dirty or faulty pH electrode
  • Temperature difference between buffer and calibration
  • Prepare fresh buffer solutions
  • Review preparation steps (Section 3)
  • Clean and recondition electrode; replace if necessary
  • Ensure buffers are at room temperature (25°C)
pH meter fails to calibrate
  • Utilizouse un búfer incorrecto
  • Electrode not immersed correctly
  • Mal funcionamento do contador
  • Verify correct buffer for calibration point
  • Ensure electrode bulb and junction are fully submerged
  • Consult pH meter manual or contact manufacturer
Cloudy or discolored buffer solution
  • Bacterial growth or contamination
  • Almacenamento inadecuado
  • Discard solution immediately
  • Prepare fresh solution using clean containers
  • Review storage guidelines (Section 5)

7. Especificacións

AtributoDetalle
Nome do produtopH Buffer Powder Set
MarcaOHAUS
Número de modelo83033971 (Item model number: 1185M39EA)
pH Values Included4.01, 7.00, 10.01
FormularioSobres de polvo
FabricanteCorporación Ohaus
ASINB01HK2DAIO
Primeiro dispoñible25 de xuño de 2016

8. Garantía e soporte

Information regarding product warranty and customer support for OHAUS products is typically provided with the product packaging or available on the official OHAUS website. For specific warranty terms, technical assistance, or to contact customer service, please visit www.ohaus.com or refer to the documentation included with your purchase.

Documentos relacionados - 83033971

Preview Manual de instrucións do analizador de humidade OHAUS MB23 / MB25
Manual de instrucións completo para os analizadores de humidade OHAUS MB23 e MB25, que abrangue a instalación, o funcionamento, a seguridade, o mantemento e a resolución de problemas para un rendemento óptimo.
Preview Manual de instrucións da interface RS232 de OHAUS para básculas Scout Pro, Traveler e Navigator
Manual de instrucións detallado para o accesorio de interface de datos RS232 de OHAUS, que abrangue a instalación, a configuración, o funcionamento, os formatos de saída e os comandos das básculas OHAUS Scout Pro, Traveler e Navigator.
Preview Ohaus Aquasearcher AB41PH/AB33PH pH Meter Quick Start Guide
A concise guide to operating the Ohaus Aquasearcher AB41PH and AB33PH pH/mV/ORP benchtop meters, covering setup, calibration, measurement, and data management.
Preview Axitadores orbitais de alta resistencia incubadores OHAUS | ISHD16HDG e ISHD23HDG: especificacións e características
Explore os axitadores orbitais de incubación de alta resistencia OHAUS (ISHD16HDG, ISHD23HDG). Obteña información sobre as características únicas, as especificacións detalladas, os accesorios dispoñibles e a información de conformidade para estes axitadores de laboratorio avanzados.
Preview Manual de instrucións do fotómetro portátil OHAUS AquaSearcher AP50MM
Manual de instrucións completo para o fotómetro portátil OHAUS AquaSearcher AP50MM, que detalla a instalación, o funcionamento e o mantemento de medicións colorimétricas e de turbidez en aplicacións de campo e de laboratorio.
Preview Manual de instrucións do axitador extremo OHAUS SHEX1619DG
Manual de instrucións completo para o axitador OHAUS Extreme, modelo SHEX1619DG. Abarca a instalación, o funcionamento, o mantemento, a seguridade, as especificacións e a resolución de problemas para uso en laboratorio.