The Okimat 6 IPS Adjustable Hospital Bed Motor Actuator is a replacement double actuator motor designed for adjustable hospital beds. It is compatible with and replaces models such as Okimat 2, 3, and 6. This unit is specifically engineered to control the back and foot sections of the bed for up and down lift functions. It comes with a wired 6-button remote hand control for easy operation.
- Compatibilidade: Replaces Okimat 2, 3, 6, and other double actuator motors.
- Funcionalidade: Controls back and foot up/down lift only; does not support massage functions.
- Control: Includes a wired 6-button remote hand control.
- Dimensións: Length x width x height: 708 mm x 118 mm x 165 mm.
- Montaxe: Features a center-to-center distance of 581mm between mounting bars.

The Okimat 6 IPS motor actuator and its accompanying 6-button wired remote hand control, designed for adjustable hospital beds.
Información de seguridade
PRECAUCIÓN: This product is designed to replace specific parts of adjustable hospital beds. Ensure it is compatible with your bed before installation. Do not attempt to replace the massage function. Use the hand control with caution to avoid tripping hazards. Always supervise children around this product.
- Always disconnect power to the bed before installation, maintenance, or troubleshooting.
- Do not operate the motor beyond its specified duty cycle (2 minutes ON / 18 minutes OFF) to prevent overheating.
- Keep all cables and electrical connections clear of moving parts and potential pinch points.
- Do not attempt to open or modify the motor housing. Refer all repairs to qualified service personnel.
Configuración e instalación
This section provides instructions for replacing your existing bed motor actuator with the Okimat 6 IPS unit. It is recommended to have assistance during installation due to the weight and size of the component.
- Preparación: Ensure the bed is unplugged from the power source before beginning any installation. Clear the area around the bed for safe access.
- Access the Existing Motor: Carefully lift or tilt the bed frame to gain access to the underside where the motor actuator is located. You may need to remove any mattress or bedding.
- Desconectar o cableado: Disconnect all electrical connections from the old motor, including the power cord and hand control cable. Note how they are routed.
- Retirar o motor antigo: Unfasten the mounting bars or brackets securing the old motor to the bed frame. Carefully remove the old motor. Note the orientation and attachment points for reinstallation. The center-to-center distance between mounting bars is 581mm.
- Posición do novo motor: Carefully position the new Okimat 6 IPS motor actuator in the same location as the old one. Ensure the mounting points align.
- Secure New Motor: Fasten the new motor securely using the appropriate mounting hardware. Verify that the motor is stable and does not wobble.
- Conectar cableado: Connect the power cord and the new 6-button remote hand control to the corresponding ports on the Okimat 6 IPS motor. Ensure connections are firm.
- Proba inicial: Plug the bed back into a power source. Before placing any weight on the bed, perform a test by using the hand control to move the head and foot sections up and down.
- Verificación final: Ensure all cables are routed safely and are not pinched or tangled by moving parts of the bed. Replace mattress and bedding.

Parte inferior view of the Okimat 6 IPS motor actuator, highlighting the mounting points for installation onto a bed frame.
Instrucións de funcionamento
The Okimat 6 IPS motor actuator is controlled by the included 6-button remote hand control. This control allows for independent adjustment of the head and foot sections of your adjustable bed.

Figura 1: Okimat 6-Button Remote Hand Control Layout.
- Head Up: Press the top-left button to raise the head section of the bed.
- Head Down: Press the bottom-left button to lower the head section of the bed.
- Foot Up: Press the top-right button to raise the foot section of the bed.
- Foot Down: Press the bottom-right button to lower the foot section of the bed.
- Simultaneous Movement: The two middle buttons operate the head and foot sections together, allowing for synchronized adjustment.
Nota importante: This motor actuator is designed to replace the back and foot up and down lift functions only. It does not support or replace any massage functions that your previous motor might have had.
Mantemento
To ensure the longevity and optimal performance of your Okimat 6 IPS motor actuator, follow these general maintenance guidelines:
- Limpeza regular: Keep the motor and hand control free from dust and debris. Use a soft, dry cloth for cleaning. Do not use liquid cleaners directly on the electrical components.
- Inspección de cables: Periodically inspect all cables for signs of wear, fraying, or damage. Ensure cables are not pinched by moving parts of the bed.
- Comprobación de conexión: Verifique que todas as conexións eléctricas estean seguras e libres de corrosión.
- Duty Cycle Adherence: The motor has a duty cycle of 2 minutes ON / 18 minutes OFF. Avoid continuous operation for more than 2 minutes to prevent overheating and potential damage. Allow the motor to rest for at least 18 minutes after 2 minutes of continuous use.
- Servizo profesional: Do not attempt to open or repair the motor housing yourself. Refer all internal repairs to qualified service personnel.
Resolución de problemas
If you encounter issues with your Okimat 6 IPS motor actuator, refer to the following common troubleshooting steps:
| Problema | Causa posible | Solución |
|---|---|---|
| Motor does not respond to hand control. |
|
|
| Bed moves slowly or with difficulty. |
|
|
| O motor fai ruídos estraños. |
|
|
Se o problema persiste despois de intentar estas solucións, póñase en contacto co servizo de atención ao cliente para obter máis axuda.
Especificacións
Detailed technical specifications for the Okimat 6 IPS Adjustable Bed Motor Actuator:
| Atributo | Detalle |
|---|---|
| Nome do modelo | Okimat 6 IPS |
| Vol. De entradatage & Frecuencia | 100 V-240 V ~ 50/60 Hz |
| Nivel de potencia | 50 W |
| Corrente máxima | 3.15A |
| Ciclo de traballo | 2 min ON / 18 min OFF |
| Max Force/Stroke (Head) | 4500N / 87mm |
| Max Force/Stroke (Foot) | 4500N / 69mm |
| Velocidade | 3.8 mm/s |
| Tipo de unidade | Dobre unidade |
| Center-to-Center Distance (Mounting Bars) | 581 mm |
| Dimensións (L x W x H) | 708 mm x 118 mm x 165 mm |
| Peso do produto | Aproximadamente 10.9 libras |

Figura 2: Technical specifications label on the Okimat 6 IPS motor.
Garantía e Soporte
While specific warranty details are not provided in this manual, Okin products are generally supported by the manufacturer. If you experience any issues with your Okimat 6 IPS Adjustable Bed Motor Actuator, or require technical assistance, please contact the seller or Okin customer support directly.
When contacting support, please have your product model number (Okimat 6 IPS) and purchase information readily available to expedite the process.
For further information or to visit the Okin store, you may refer to the official brand page: Tenda Okin en Amazon.