U.S. Solid 5000g x 0.01g Analytical Balance

U.S. Solid Precision Lab Scale 5000g x 0.01g Analytical Balance Instruction Manual

Model: 5000g x 0.01g Analytical Balance with RS232 Interface

1. Introdución

This manual provides comprehensive instructions for the safe and efficient operation of your U.S. Solid Precision Lab Scale. Please read this manual thoroughly before use and retain it for future reference. This analytical balance is designed for precise weighing in laboratory, industrial, and educational settings, offering a capacity of 5000g with 0.01g readability.

2. Precaucións de seguridade

  • Always operate the scale on a stable, level, and vibration-free surface.
  • Evite expoñer a báscula a temperaturas extremas, humidade ou luz solar directa.
  • Non sobrecargue a báscula máis alá da súa capacidade máxima de 5000 g.
  • Keep the weighing pan clean and free from debris.
  • Use só o adaptador de alimentación proporcionado.
  • Avoid sudden impacts or dropping the scale.

3. Configuración e instalación

3.1 Desembalaxe

Carefully remove the scale and all accessories from the packaging. Verify that all components listed in the packing list are present. The package should include the scale, stainless steel weighing pan, and power adapter.

3.2 Colocación

Place the scale on a firm, level surface away from drafts, vibrations, and direct heat sources. Ensure adequate space around the scale for operation and ventilation.

3.3 Nivelación da escala

The scale must be perfectly level for accurate measurements. Use the adjustable feet at the bottom of the scale and the bubble level indicator to achieve proper leveling.

U.S. Solid Precision Lab Scale with bubble level indicator

Imaxe: Fronte view of the U.S. Solid Precision Lab Scale, showing the bubble level indicator on the left side of the display panel.

Diagram showing horizontal adjustment for leveling the scale

Image: Diagram illustrating how to adjust the leveling feet until the bubble is centered, ensuring proper horizontal adjustment for accuracy.

3.4 Power Connection and Warm-up

Connect the provided power adapter to the scale and then to a suitable power outlet. For optimal accuracy, allow the scale to warm up for at least 30 minutes after connecting to power or after a power loss before performing any measurements.

Video: This video demonstrates the initial setup of the U.S. Solid Precision Lab Scale, including connecting the power supply and the recommended 30-minute warm-up period for optimal performance.

4. Instrucións de funcionamento

4.1 Pesaxe básica

  1. Preme o ON/OFF botón para encender a báscula.
  2. Wait for the display to show "0.00g".
  3. Place the item to be weighed on the center of the stainless steel pan.
  4. Lea o peso estable que se mostra na pantalla LCD.

4.2 Conversión de unidades

The scale supports 19 measurement units. Press the UNIDADE button to cycle through the available units (g, kg, ct, T, TAR, dr, PKT, GN, TMR, gsm, tlJ, mo, dwt, oz, lb, tlT, ozt, tlH, %).

Diagram showing 19 measurement units available on the scale

Image: Visual representation of the 19 different measurement units supported by the U.S. Solid Precision Lab Scale, including grams, kilograms, carats, and more.

4.3 Función de tara

To subtract the weight of a container, place the empty container on the pan and press the TARE/CAL button. The display will reset to "0.00g". You can then add the item to be weighed into the container, and only the net weight of the item will be displayed.

4.4 Piece Counting Function

The scale can count pieces of uniform weight. To use this function:

  1. Colocar comoample quantity (e.g., 10, 20, 50, 100, 150, 200, 250, or 500 pieces) of the items to be counted on the pan.
  2. Manteña premido o COU button until "PCS" appears on the display.
  3. Usa o UNIDADE botón para seleccionar o sample quantity that matches the number of items on the pan.
  4. Preme TARE/CAL to confirm. The display will now show the number of pieces.
  5. Add more items to the pan, and the scale will display the total count.
Image showing piece counting function with various sampos tamaños

Image: Illustration of the piece counting feature, demonstrating how to set sample sizes (10, 20, 50, etc.) for accurate item counting.

4.5 Configuración de retroiluminación

To adjust the backlight behavior:

  1. Preme o COU button to choose between C0-OFF, C0-1, or C0-5.
  2. C0-OFF: Backlight is always on.
  3. C0-1: Backlight will turn off after 1 second of inactivity.
  4. C0-5: Backlight will turn off after 5 seconds of inactivity.
  5. Finalmente, fai clic TAREAR until the program ends and returns to 0.00.

Video: This segment demonstrates how to adjust the backlight settings of the scale, allowing users to choose between always-on, 1-second, or 5-second inactivity timeouts for the backlight.

4.6 Unit Package Locking

To lock specific unit packages for quick access:

  1. Manteña premido o COU button. Release it when "C0-OFF" is displayed.
  2. Fai clic TAREAR to choose C06-1.
  3. Preme o COU button to cycle through C06-1, C06-2, up to C06-16, each representing a different combination of units.
  4. Finalmente, fai clic TAREAR until the program ends and returns to 0.00.

Video: This segment illustrates how to lock specific unit packages on the scale, allowing users to customize and quickly switch between preferred sets of measurement units.

5. RS232 Interface Communication

The U.S. Solid Precision Lab Scale features an RS232 interface for connecting to a computer or printer, enabling serial data exchange. This allows for data logging and integration with other systems.

Lado view of the U.S. Solid Precision Lab Scale showing the RS232 port

Imaxe: Lateral view of the U.S. Solid Precision Lab Scale, highlighting the RS232 interface port for external connectivity.

5.1 Key Output Mode Setting (C5-2)

In Key Output Mode, the scale sends weighing data to the connected device only when the UNIDADE preme o botón.

  1. Manteña premido o COU button. Release it when "C0-OFF" is displayed.
  2. Fai clic TAREAR to choose C5-0.
  3. Preme o COU button to choose C5-2.
  4. Finalmente, prema o TAREAR button to enter into key output mode.
  5. To output data, open "Uartassist" software (or similar serial terminal software), set Serial Port to COM3, Baud Rate to 9600, Data Bits to 8, Parity to None, and Stop Bits to One. Click "Open serial port".
  6. Cada vez que prema o botón UNIDADE button on the scale, the computer will output the current weighing data.

Video: This segment demonstrates how to set the scale to "Key Output" mode (C5-2) for USB interface testing. It shows connecting the scale to a computer and manually triggering data output by pressing the UNIT button.

5.2 Continuous Output Mode Setting (C5-0)

In Continuous Output Mode, the scale continuously sends weighing data to the connected device without requiring manual input.

  1. Manteña premido o COU button. Release it when "C0-OFF" is displayed.
  2. Fai clic TAREAR to choose C5-0.
  3. Finalmente, prema o TAREAR button to enter into continuous output mode.
  4. With "Uartassist" software (or similar serial terminal software) open and configured (COM3, 9600 baud, 8 data bits, no parity, one stop bit), click "Open serial port". The weighing results will be continuously output on the computer.

Video: This segment demonstrates how to set the scale to "Continuous Output" mode (C5-0) for USB interface testing. It shows the scale continuously sending weighing data to a connected computer.

6. Calibración

Regular calibration ensures the accuracy of your scale. The scale supports manual calibration.

  1. Ensure the scale is on a stable surface and has completed its warm-up period.
  2. With the scale displaying "0.00g", press and hold the TARE/CAL button until "CAL" appears.
  3. A pantalla mostrará entón un valor de peso intermitente (por exemplo, "2000 g"). Este é o peso de calibración requirido.
  4. Carefully place the specified calibration weight onto the center of the pan.
  5. Wait for the display to stabilize and show "PASS" or return to "0.00g".
  6. Remove the calibration weight. The scale is now calibrated.
Diagram showing calibration process for the scale

Image: Diagram illustrating the calibration process, emphasizing the need to recalibrate after moving the balance or before use, and to hold the TARE/CAL button for instructions.

7. Mantemento e coidados

  • Limpeza: Wipe the scale's exterior with a soft, damp cloth. Do not use abrasive cleaners or immerse the scale in water. The stainless steel pan can be removed for separate cleaning.
  • Almacenamento: When not in use for extended periods, store the scale in a dry, dust-free environment.
  • Inspección: Comprobe periodicamente o cable de alimentación para ver se está danado.

8 Solución de problemas

ProblemaCausa posibleSolución
A escala non se acendeNo power, faulty adapter, or power outlet issue.Check power connection, try a different outlet, ensure adapter is functional.
Lecturas inexactasNot level, drafts, vibrations, not warmed up, needs calibration, or overloaded.Level the scale, move to a stable location, allow warm-up, perform calibration, ensure weight is within capacity.
Display shows "O-Ld"Sobrecarga.Retire o exceso de peso inmediatamente para evitar danos.
Fallo de comunicación RS232Incorrect port settings, cable issue, or wrong output mode.Verify COM port, baud rate, data bits, parity, and stop bits. Check cable connection. Ensure correct output mode (Key or Continuous) is selected.

9. Especificacións

  • Marca: Sólido estadounidense
  • Modelo: 5000g x 0.01g Analytical Balance
  • Capacidade: 5000 gramos (5 kg)
  • Lexibilidade: 0.01 Grams (10 mg)
  • Tipo de visualización: LCD con retroiluminación
  • Interface: RS232
  • Unidades de medida: 19 units (g, kg, ct, T, TAR, dr, PKT, GN, TMR, gsm, tlJ, mo, dwt, oz, lb, tlT, ozt, tlH, %)
  • Aplicacións: Check-weighing, Tare, Piece Counting, Manual Calibration, Overload Protection
  • Material: Carcasa de ABS e bandexa de aceiro inoxidable
  • Dimensións do produto: 8.9 x 8.7 x 3.2 polgadas
  • Peso do artigo: 2.9 libras
  • Potencia: Adaptador de CA (incluído)

10. Garantía e soporte

For warranty information and technical support, please refer to the warranty card included with your product or visit the official U.S. Solid websitio. Garde o recibo da compra como xustificante da mesma.

Información de contacto: Consulte a información oficial do fabricante website or contact details provided in the product packaging for support.

Documentos relacionados - 5000g x 0.01g Analytical Balance

Preview U.S. Solid USS-DBS Analytical Balance Operating Manual
Comprehensive operating manual for the U.S. Solid USS-DBS series analytical balance, covering features, specifications, installation, usage, maintenance, and troubleshooting.
Preview U.S. Solid USS-DBS82 Precision Balance Instruction Manual
Comprehensive instruction manual for the U.S. Solid USS-DBS82 Series Precision Balance, covering safety precautions, design, operation, technical specifications, and contact information.
Preview U.S. Solid USS-DBS Analytical Balance Operating Manual
Comprehensive operating manual for the U.S. Solid USS-DBS Analytical Balance, detailing specifications, installation, operation, internal settings, maintenance, troubleshooting, and legal use restrictions.
Preview U.S. Solid USS-DBS Analytical Balance Operating Manual
Comprehensive operating manual for the U.S. Solid USS-DBS series analytical balances, covering features, specifications, installation, usage, calibration, maintenance, troubleshooting, and RS232 interface.
Preview U.S. SOLID Analytical Balance Manual: Precision Weighing, Density & Dynamic Functions
Comprehensive instruction manual for U.S. SOLID 0.1 mg Analytical Balances, covering models like USS-DBS62 and USS-DBS63. Learn about precision weighing, density determination, dynamic weighing, calibration, operation, and troubleshooting for laboratory and industrial applications.
Preview U.S. Solid USS-DBS Digital Balance Operating Manual
Comprehensive operating manual for U.S. Solid USS-DBS series digital balances, covering specifications, installation, usage, maintenance, and troubleshooting. Learn how to calibrate, tare, and use piece counting functions.