National Instruments NI-9218 Channel Analogue Input Module
Especificacións
- Nome do produto: NI-9218
- Tipos de conectores: LEMO e DSUB
- Tipos de medición: Compatibilidade integrada con varios tipos
- Excitación do sensor: excitación opcional de 12 V
Tipos de conectores
O NI-9218 ten máis dun tipo de conector: NI-9218 con LEMO e NI-9218 con DSUB. A menos que se especifique o tipo de conector, NI-9218 refírese a ambos os tipos de conectores.
Desconexión de pines do NI-9218
Signals by Measurement Type
Modo | Pin
1 |
|||||||||
1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | |
± 16 V | EX+ | — | IA-, EX- | — | — | AI+ | — | — | — | — |
±65 mV | EX+ 2 [2] | — | EX- [2] | — | — | AI+ | AI- 3 | — | — | — |
completo-
Ponte |
EX+ [2] | — | EX- [2] | RS+ | RS- | AI+ | AI- | SC | SC | — |
IEPE | — | AI+ | AI- | — | — | — | — | — | — | — |
TEDS | — | T+ 4 | T- | — | — | — | — | — | — | T+ 5 |
Descricións de sinal
Sinal | Descrición |
AI+ | Positive analogue input signal connection |
AI- | Negative analogue input signal connection |
EX+ | Positive sensor excitation connection |
EX- | Negative sensor excitation connection |
RS+ | Positive remote sensing connection |
RS- | Negative remote sensing connection |
SC | Shunt calibration connection |
T+ | TEDS data connection |
T- | TEDS return connection |
Tipos de Medición
O NI-9218 ofrece compatibilidade integrada cos seguintes tipos de medición.
- ± 16 V
- ±65 mV
- Full-Bridge
- IEPE
- NI-9218 with LEMO only.
- Optional sensor excitation.
- Tie to pin 3.
- TEDS Class 1 data connection.
- TEDS Class 2 data connection.
Consello NI recommends using the NI-9982 screw-terminal adapter when using built-in measurement types on the NI-9218.
O NI-9218 ofrece compatibilidade adicional cos seguintes tipos de medición ao usar un adaptador específico para a medición.
- ±20 mA, requires the NI-9983
- ±60 V, requires the NI-9987
- Half-Bridge requires the NI-9986
- Quarter-Bridge requires the NI-9984 (120 Ω) or NI-9985 (350 Ω)
Conexións de ±16 V
O NI-9218 proporciona excitación de sensor de 12 V opcional. Para usar a excitación de 12 V, conecte unha fonte de alimentación de 9 V CC a 30 V CC a Vsup, conecte os terminais de excitación do sensor a EX+/EX- e active a excitación de 12 V no software.
Referencia relacionada:
- Conexión de ±16 V do NI-9982
Conexións de ±65 mV
- You must connect AI to EX- on the NI-9218.
- O NI-9218 proporciona excitación de sensor de 12 V opcional. Para usar a excitación de 12 V, conecte unha fonte de alimentación de 9 V CC a 30 V CC a Vsup, conecte os terminais de excitación do sensor a EX+/EX- e active a excitación de 12 V no software.
Referencia relacionada
- Conexión NI-9982 ±65 mV
Conexións de ponte completa
- The NI-9218 provides 2 V excitation to loads ≥120 Ω or 3.3 V excitation to loads ≥350 Ω.
- The NI-9218 provides optional connections for remote sensing (RS) and shunt calibration (SC). Remote sensing corrects for errors in excitation leads, and shunt calibration corrects for errors caused by resistance within one leg of the bridge.
Referencia relacionada:
- Deseño de pines da conexión de ponte completa NI-9982
Conexións do IEPE
- The NI-9218 provides an excitation current for each channel that powers IEPE sensors.
- AI+ provides DC excitation, and AI- provides the excitation return path.
Referencia relacionada:
- Pinout de conexión NI-9982 IEPE
Conexións de ±20 mA
- Connecting ±20 mA signals requires the NI-9983.
- O NI-9218 proporciona excitación de sensor de 12 V opcional. Para usar a excitación de 12 V, conecte unha fonte de alimentación de 9 V CC a 30 V CC a Vsup, conecte os terminais de excitación do sensor a EX+/EX- e active a excitación de 12 V no software.
Connecting a loop-powered 2-wire or 3-wire transducer requires adding a 20 kΩ resistor between AI- and Ex-.
Referencia relacionada:
- Desconexión de pines do NI-9983
Conexións de ±60 V
Connecting ±60 V signals requires the NI-9987.
Referencia relacionada:
- Desconexión de pines do NI-9987
Conexións de media ponte
- Connecting half bridges requires the NI-9986.
- The NI-9218 provides 2 V excitation to half bridges of ≥240 Ω total or 3.3 V excitation to half bridges of ≥700 Ω total.
- The NI-9218 provides optional connections for remote sensing (RS) and shunt calibration (SC). Remote sensing corrects for errors in excitation leads, and shunt calibration corrects for errors caused by resistance within one leg of the bridge.
Referencia relacionada:
- Desconexión de pines do NI-9986
Conexións de cuarto de ponte
- Connecting 120 Ω quarter bridges requires the NI-9984.
- Connecting 350 Ω quarter bridges requires the NI-9985.
Consello NI-recommends 2 V excitation when using a NI-9984 with 120 Ω quarter bridges and 3.3 V excitation when using the NI-9985 with 350 Ω quarter bridges.
Referencia relacionada:
- Desconexión de pines NI-9984/9985
Conexións TEDS
For more information about TEDS, visit ni.com/info and enter the Info Code rdteds.
Soporte TEDS
- Os sensores TEDS de clase 1 proporcionan unha interface para transferir información desde os sensores. Os modelos NI-9218 con LEMO, NI-9218 con DSUB, NI-9982L, NI-9982D e NI-9982F admiten sensores TEDS de clase 1.
- TEDS Class 2 sensors provide an interface for transferring information from TEDS-enabled sensors. The NI-9218 with LEMO, NI-9982L, NI-9983L, NI-9984L, NI-9985L, and NI-9986L support TEDS Class 2 sensors.
Topoloxía de cadea de margaridas Vsup
O NI-9218 con LEMO proporciona catro pines no conector Vsup para a conexión en cadea.
Directrices de conexión NI-9218
Make sure that the devices you connect to the NI-9218 are compatible with the module specifications.
Directrices de cableado personalizado
- Siga as seguintes pautas ao usar o adaptador de conector de copa de soldadura NI-9988 ou o conector de engarzado LEMO (784162-01) para crear cables personalizados.
- Use a shielded cable for all signals.
- Connect the cable shield to the earth ground.
- Use twisted-pair wiring for the AI+/AI- and RS+/RS- signals to achieve specified EMC performance.
Diagrama de bloques NI-9218
- Two 24-bit analogue-to-digital converters (ADCs) simultaneously sampambos os canais de IA.
- The NI-9218 provides channel-to-channel isolation.
- The NI-9218 reconfigures the signal conditioning for each measurement type.
- The NI-9218 provides excitation for IEPE and bridge completion measurement types.
- The NI-9218 can provide optional 12 V sensor excitation for ±16 V, ±65 mV, and ±20 mA measurement types.
Acondicionamento de sinal de ±16 V e ±65 mV
Os sinais de entrada en cada canal almacénanse en búfer, acondiciónanse e logo emítenseampdirixido por un ADC.
Acondicionamento de sinal de ponte completa
- The analogue input connections sense and then amplify the incoming analogue signal.
- The excitation connections provide differential bridge-excitation voltage.
- Remote sensing continuously and automatically corrects for lead-wire induced excitation voltage loss when using the RS connections.
- Shunt calibration can be used to correct for lead-wire-induced desensitisation of the bridge.
Acondicionamento de sinais IEPE
- The incoming analogue signal is referenced to an isolated ground.
- Each channel is configured for AC coupling with an IEPE current.
- Each channel provides a TEDS Class 1 interface.
Acondicionamento de sinal de ±20 mA
The NI-9983 provides a current shunt for the incoming analogue signal.
Acondicionamento de sinal de ±60 V
The NI-9987 provides an attenuator for the incoming analogue signal.
Acondicionamento de sinal de media ponte
- The NI-9886 provides half-bridge completion resistors for the incoming analogue signal.
- You must connect AI+, EX+, and EX-.
- RS+ and RS- connections are optional.
- You do not need to connect the AI signal because it is connected internally.
Acondicionamento do modo de cuarto de ponte
The NI-9984 and NI-9985 provide a quarter-bridge completion resistor and a half-bridge completion resistor.
Filtrado
The NI-9218 uses a combination of analogue and digital filtering to provide an accurate representation of in-band signals while rejecting out-of-band signals. The filters discriminate between signals based on the frequency range, or bandwidth, of the signal. The three important bandwidths to consider are the passband, the stopband, and the alias-free bandwidth.
O NI-9218 representa os sinais dentro da banda de paso, cuantificados principalmente pola ondulación da banda de paso e a non linealidade de fase. Todos os sinais que aparecen na banda de paso libre de alias son sinais sen alias ou sinais que foron filtrados polo menos na mesma cantidade que o rexeitamento da banda de parada.
Banda de paso
Os sinais dentro da banda de paso teñen unha ganancia ou atenuación dependente da frecuencia. A pequena cantidade de variación na ganancia con respecto á frecuencia chámase planitude da banda de paso. Os filtros dixitais do NI-9218 axustan o rango de frecuencias da banda de paso para que coincida coa taxa de datos. Polo tanto, a cantidade de ganancia ou atenuación a unha frecuencia determinada depende da taxa de datos.
Banda de parada
O filtro atenúa significativamente todos os sinais por riba da frecuencia da banda de parada. O obxectivo principal do filtro é evitar o aliasing. Polo tanto, a frecuencia da banda de parada escálase con precisión coa taxa de datos. O rexeitamento da banda de parada é a cantidade mínima de atenuación aplicada polo filtro a todos os sinais con frecuencias dentro da banda de parada.
Ancho de banda sen alias
Any signal that appears in the alias-free bandwidth of the NI-9218 is not an aliased artefact of signals at a higher frequency. The alias-free bandwidth is defined by the ability of the filter to reject frequencies above the stopband frequency, and it is equal to the data rate minus the stopband frequency.
Abrir un adaptador de medición
Que facer
- Unlock the measurement adapter housing/cover.
- Slide the measurement adapter housing/cover to access the screw terminals.
Montaxe do NI-998xD/998xL
Que usar
- NI-998xD or NI-998xL Measurement Adapter
- M4 or Number 8 Screw
- Desaparafusador
Que facer
Mount the measurement adapter to a flat surface using the mounting hole on the measurement adapter and the screw.
Adaptador de medición de conexión a terra
Os terminais de terra dun adaptador de medición están conectados á terra do chasis cando o adaptador de medición está conectado ao NI-9218 e o NI-9218 está instalado nun chasis.
Pinouts do adaptador de medición
As seguintes seccións inclúen as conexións dos adaptadores de medición NI-9218.
Conexión de ±16 V do NI-9982
Pins 3a and 3b are tied together on the NI-9982.
Referencia relacionada:
- Conexións de ±16 V
Conexión NI-9982 ±65 mV
Pins 3a and 3b are tied together on the NI-9982.
Referencia relacionada:
- Conexións de ±65 mV
Deseño de pines da conexión de ponte completa NI-9982
Pins 3a and 3b are tied together on the NI-9982.
Referencia relacionada:
- Conexións de ponte completa
Pinout de conexión NI-9982 IEPE
Pins 3a and 3b are tied together on the NI-9982.
Referencia relacionada:
- Conexións do IEPE
Desconexión de pines do NI-9983
Pins 3a and 3b are tied together on the NI-9983.
Referencia relacionada:
- Conexións de ±20 mA
Desconexión de pines NI-9984/9985
Referencia relacionada:
- Conexións de cuarto de ponte
Desconexión de pines do NI-9986
Pins 3a and 3b are tied together on the NI-9986.
Referencia relacionada:
- Conexións de media ponte
Desconexión de pines do NI-9987
Pins 3a and 3b are tied together on the NI-9987.
Referencia relacionada:
- Conexións de ±60 V
FAQ
Documentos/Recursos
![]() |
Módulo de entrada analóxica de canle NI-9218 de National Instruments [pdfManual de instrucións NI-9218 con LEMO, NI-9218 con DSUB, Módulo de entrada analóxica de canle NI-9218, NI-9218, Módulo de entrada analóxica de canle, Módulo de entrada analóxica, Módulo de entrada |